1. Home
  2. Knowledge Base
  3. BITalino
  4. BITalino Electrodermal Activity (EDA) Signal Troubleshooting
  1. Home
  2. Knowledge Base
  3. Sensors
  4. BITalino Electrodermal Activity (EDA) Signal Troubleshooting
  1. Home
  2. Knowledge Base
  3. Hardware
  4. BITalino Electrodermal Activity (EDA) Signal Troubleshooting

BITalino Electrodermal Activity (EDA) Signal Troubleshooting

Before starting an Electrodermal Ectivity (EDA) measurement with the BITalino sensor we recommend you to check if you are following all our recommendations (below).

Skin cleaning

The first thing you should do when planning to acquire an EDA signal is to clean the skin surface, remove skin debris, dust, or oil that is present on the skin surface, and might interfere with the conductivity measurements. You can use an alcohol-based solution at 96% (v/v) to clean your skin before placing the electrodes.

Electrodes

EDA measurements are very sensitive to changes in electrode-to-skin conductivity. So never re-use electrodes.

Try to be consistent in using electrodes, i.e., if you start an experiment with pre-gelled and dischargeable electrodes, keep the same as reusable ones during your set of experiments (unless the goal is to compare the electrodes’ performance). You might notice differences in the values depending on the type of electrodes you’re using.

Electrodes positioning

The BITalino EDA sensor is designed to acquire the change of skin activity such as sweat with two measuring electrodes. One example is the placement of the electrodes on the anterior side of the hand on two adjacent fingers of interest (see Fig. 1). The electrodes are then connected to the sensor cable of the EDA sensor.

Figure 1: Example electrode placement of the assembled version on the anterior side of the hand on the index and middle finger (left) and below the thumb (right).

Sensors set-up in BITalino

When using BITalino we do not recommend the use of more than one EDA sensor at the same time.

Due to the technical design of the EDA sensor of the BITalino, you may notice some irregularities in this signal when working with it together with other ExG sensors (e.g., electrocardiography (ECG), electromyography (EMG), electroencephalography (EEG), etc.) that require the use of a reference cable (Vref).

When you attach the Vref of the ExG sensor, you will have a reference at two different points on the body, once the EDA sensor has its own intrinsic reference as well. These two reference points might interfere in the EDA resistance measurements, which will drop significantly, while acquiring the signals at the same time.

To overcome this issue, we recommend using ExG sensors also with 2-lead electrode cables. You can also use with 3-lead electrode, then you must disconnect the midpoint of the ExG sensor (i.e. the Vref, usually labeled with white sleave) from the body and leaving the electrode in the air. This procedure has no effect on the ExG signals measurement.

Updated on 19 de December de 2022

Was this article helpful?

Related Articles

Need Support?
Get one-to-one support for your Plux device or sensor from our team of biosignals experts.
Get Support